Comparison of Different Methods for Measurement of Intraocular Pressure and Evaluation of Effect of Central Corneal Thickness on Measurement Results

Turk J Ophthalmol 2010;40(3):156-159
1.
2.
No information available.
No information available
Received Date: 23.03.2010
Accepted Date: 14.10.2009
PDF
Cite
Share
Request

ABSTRACT

Discussion:

There was a significant difference between measurements of IOP by GAT, NCT and tonopen in normal eyes. Tonopen may not be affected by CCT in normal eyes, while a significant correlation existed between CCT and GAT or NCT in normal eyes.

Results:

The mean IOP measured by GAT was 12.63±2.94 mmHg, by tonopen - 8.51±2.66 mmHg and by NCT - 13.34±3.88 mmHg. In terms of measurement results, there was a statistically significant difference between the three methods (p<0.05). Although the correlation between tonopen and CCT was insignificant (p>0.05), there was a significant correlation between GAT and CCT, as well as between NCT and CCT (p<0.05).

Material and Method:

Seventy-nine eyes of 41 patients, whose ophthalmologic examination and visual field test revealed no glaucoma, were included into the study. IOP was measured by GAT, NCT and tonopen. CCT was measured with an ultrasonic pachymeter.

Purpose:

To compare measurements of intraocular pressure (IOP) by Goldmann applanation tonometer (GAT), tonopen and non-contact tonometer (NCT) in normal eyes, and determine the relationship between the results of IOP measurement and central corneal thickness (CCT).