Dear Editor,
The recent publication on “Pediatric Patients and Tonometers” is interesting.1Eraslan et al.1concluded that “Because Tono-Pen (TP) measurements were lower than Goldmann applanation tonometer (GAT) measurements and non-contact tonometer measurements were higher than GAT measurements, patient follow-ups, treatment strategies, and surgery plans must be organized taking these differences into consideration”. The results in this report are similar to a recent report by Galgauskas et al.2In fact, the use of different kinds of tonometer can result in different measures values and this has to be kept in mind by practitioners. The correlation study can be useful for checking the variability of the tool. Nevertheless, there are some concerns that should be addressed. First, the lack of a gold standard for the comparative study is a big issue for further discussion. At present, we can only perform the inter-tool agreement check but there is no gold standard for checking the accuracy of the measurement. Second, for each tool the within-day and between-day precision of the tool should also be checked. Finally, the calibration error of the tool should be regularly checked since it can contribute to incorrect measurement results.3 In the present report by Eraslan et al.,1 there is no error checking as well.
Keywords:
Pediatric, Tonometer, measurement
Ethics
Peer-review: Internally peer-reviewed.
Authorship Contributions
Surgical and Medical Practices: Sora Yasri, Viroj Wiwanitkit, Concept: Sora Yasri, Viroj Wiwanitkit, Design: Sora Yasri, Viroj Wiwanitkit, Data Collection or Processing: Sora Yasri, Viroj Wiwanitkit, Analysis or Interpretation: Sora Yasri, Viroj Wiwanitkit, Literature Search: Sora Yasri, Viroj Wiwanitkit, Writing: Sora Yasri, Viroj Wiwanitkit.
Conflict of Interest: No conflict of interest was declared by the authors.
Financial Disclosure: The authors declared that this study received no financial support.
1. Eraslan M, Eren Çerman E, Sümmen S. Comparison of Intraocular Pressure Measurements in Healthy Pediatric Patients using Three Types of Tonometers. Turk J Ophthalmol. 2017;47:1–4.
PubMed]2. Galgauskas S, Strupaite R, Strelkauskaite E, Asoklis R. Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons. Int J Ophthalmol. 2016;9:76–80. [
PubMed]3. Choudhari NS, Rao HL, Ramavath S, Rekha G, Rao A, Senthil S, Garudadri CS. How Often the Goldmann Applanation Tonometer Should be Checked for Calibration Error? J Glaucoma. 2016;25:908–913. [
PubMed]
“Pediatrik Hastalar ve Tonometreler” ile ilgili makaleyi çok ilgi çekici bulduk.1Eraslan ve ark.1Tono-Pen (TP) ölçümleri Goldmann aplanasyon tonometre (GAT) ölçümlerinden daha düşük sonuç verdiği ve non-kontakt tonometre ölçümleri ise GAT ölçümlerinden daha yüksek sonuç verdiği için hastaların izleminde, tedavi stratejisi belirlenirken ve cerrahi planlanırken bu farkların göz önüne bulundurulması gerektiği sonucuna varmıştır. Bu çalışmanın sonuçları yakın zamanda Galgauskas ve ark.2tarafından yapılan çalışma ile benzerdir. Gerçekten de farklı tonometre tiplerinin kullanılmasının değişik ölçüm değerlerinin elde edilmesine neden olabileceği akılda tutulmalıdır. Bu aletlerin değişkenliğinin araştırılmasında korelasyon çalışmaları yararlı olabilir. Ancak, açıklığa kavuşturulması gereken noktalar bulunmaktadır. İlk olarak, karşılaştırmalı bir çalışma için altın standardın bulunmaması daha ileri tartışma gerektiren önemli bir eksiktir. Şu anda, sadece farklı aletler ile yapılan ölçümleri birbiri ile karşılaştırabiliyoruz ancak ölçümün doğruluğunu kontrol etmek için bir altın standart mevcut değil. İkinci olarak, her aletin gün içinde ve günler arası hassasiyeti kontrol edilmelidir. Son olarak, aletin kalibrasyon hatası düzenli olarak kontrol edilmelidir çünkü, bu durum hatalı ölçüm sonuçlarına katkıda bulunabilir.3Eraslan ve ark.1tarafından yapılan çalışmada hata kontrolü de yapılmamıştır.Etik
Hakem Değerlendirmesi: Editörler kurulu tarafından değerlendirilmiştir.
Yazarlık Katkıları
Cerrahi ve Medikal Uygulama: Sora Yasri, Viroj Wiwanitkit, Konsept: Sora Yasri, Viroj Wiwanitkit, Dizayn: Sora Yasri, Viroj Wiwanitkit, Veri Toplama veya İşleme: Sora Yasri, Viroj Wiwanitkit, Analiz veya Yorumlama: Sora Yasri, Viroj Wiwanitkit, Literatür Arama: Sora Yasri, Viroj Wiwanitkit, Yazan: Sora Yasri, Viroj Wiwanitkit.
Çıkar Çatışması: Yazarlar tarafından çıkar çatışması bildirilmemiştir.
Finansal Destek: Yazarlar tarafından finansal destek almadıkları bildirilmiştir.
We are grateful to Yasri and Wiwanitkit for their interest in our work and their valuable comments. The authors kindly reminded that the use of different kinds of tonometer can result in different measured values, which corresponds to the conclusion of our study. They mentioned the lack of a gold standard in intra ocular pressure measurements but as all we know, the Goldmann applanation tonometer (GAT) is defined as the gold standard measurement method in a large number of studies.1 As they emphasized in their letter, the findings of our study are similar to a recent report of Galgauskas et al.2 where GAT is defined as the gold standard. But as Garcia Feijoo et al.3 mentioned in their study, despite GAT being the gold standard for determining intra ocular pressure since the last century, the substantial effects of several ocular variables such as axial length, curvature, rigidity, and corneal thickness are its obvious limitations. Yasri and Wiwanitkit also mentioned that we can only perform the inter-tool agreement and they recommended that a correlation study can be useful for checking the variability of the tool. However, to evaluate whether the differences between two measurements of the same variable are significant, previous studies recommend studying the differences, not the agreement.4 The correlation shows the relationship between one variable and another, not the differences, and it is not the best technique for assessing the comparability between methods.4 Bland-Altman (B-A) plots compare two clinical measurements that each provide some errors in their measure and these plots are extensively used to evaluate the agreement between two different instruments or two measurement techniques.5 B-A plot analysis can also be used for assessing the comparability between a new measurement technique or method with a gold standard, as even a gold standard does not-and should not-imply it to be without error.6 These analyses evaluate a bias between the mean differences and estimate an agreement interval, within which fall 95% of the differences of the second method compared to the first one. It is common to compute 95% limits of agreement for each comparison (average difference ±1.96 standard deviation (SD) of the difference). The compared methods can be used interchangeably unless the differences within mean ±1.96 SD are clinically important.4
In our study, Pearson’s test was used to determine the presence of correlations. Differences of 1.96 SD from the mean were used when calculating the limits of agreement. Associations between differences and means were analyzed using B-A plots. This was mentioned in the third paragraph of the Materials and Methods section of our study.
Within-day and between-day precision was checked for each tool and because our GAT and Tonopen are older than 1 year, the calibration error of the tools is routinely checked on a daily basis and the non-contact tonometer was calibrated once a month as recommended by the manufacturers. These were also mentioned in the second paragraph of the Materials and Methods section of our study. This is consistent with the study of Choudhari et al.7, which was mentioned by Yasri and Wiwanitkit in their letter; they concluded that GATs older than a year should be checked at least monthly. Therefore, we believe that all of the abovementioned limitations are important but it is unlikely that they affected our results significantly.
Muhsin Eraslan, Eren Çerman, Sena Sümmen